Moisture  Meter Calibration Plate

Moisture Meter Calibration Plate

gray stars

Click Picture for more Info

Calibration plate for CT33, Dolphin and Pinless Meters.

Can$49.99
SKU 13015
Qty

Calibration plate forCT33, Dolphin and Pinless Meters Keep one of these low cost plate on hand for meter calibration.

ZERO AND CALIBRATION INSTRUCTIONS FOR THE DOLPHIN 

FIRST CHECK THE ZERO-POINT

Switch on the meter and hold it horizontally in mid-air, with your finders well away from the sensing area on the rear. Ther meter should read 0%.

ADJUST THE ZERO IF NECESSARY AS FOLLOWS

On the front panel there are two pale, almost white, plastic screw heads. Remove the left-hand screw, to expose a small hole in the meter body. Through the hole will be visible a white circle with a slot, just like a slotted screw head. Turn on the meter and hold it horizontally in mid-air. Take a small flat-bladed screwdriver and carefully insert the screwdriver blade through the hole to engage the white slot. Turn the screw very slowly until the needle points to the 0% mark. Re-insert the white plastic screw cap now to avoid later confusion.

CALIBRATION CHECK (USING OPTIONAL ACCESSORY CALIBRATION PLATE)

There are two plates. Take the opaque copper-coated plate and place on a table top or other convenient non-metallic surface, with the copper side facing down.

Place the second (clear plastic) plate on top of the first.

Turn on the meter and place directly on top of the clear plastic plate and press gently. The meter should read 14%. If it doesn't proceed as follows.

CALIBRATION ADJUSTMENT

The meter remains on top of the clear plastic plate on top of the opaque plate with copper side down.

Remove the right-hand plastic screw to expose a small hole in the meter body. Through the hole is visible a white circle with a slot, just like a head of a screw. Take a small flat-bladed screwdriver and carefully insert the blade into the exposed hole to engage the slot. Turn the screw very slowly until the meter reads 14%. Re-insert the white plastic screw to keep foreign materials out of the meter.

Both calibration plates should be kept flat, clean and smooth. The copper-coated plate should remain untarnished and corrosion free.


 Shipping Chemicals and  Wide Fabric Make it impossible to Calculate Shipping Costs in Advance.
Your trust is important so we only charge what shipping cost us.
 
Disclaimer: The information contained herein and provided by RAYPLEX LTD is considered accurate.
However, no warranty is expressed or implied regarding the accuracy of the data, the results to be obtained
by the use thereof, or that any such use will not infringe any patent. Before using, user shall determine
the suitability of the product for the intended use and user assumes all risk and liability whatsoever
in connection therewith.
 

CUSTOMER SERVICES

WEBSITE INFORMATION
RESOURCES

RAYPLEX LTD

Returns
Shipping 
Indian Status Tax Exemption

 

 

 

 

 
 

 

 

 
Copyright © 2024 Rayplex Ltd.  
RAYPLEX is a registered
Trade Mark All rights reserved. 
 
 
 
 
341 Durham Crt, Oshawa, ON L1J 1W8
Phone: 905-579-1433 Fax: 905-579-1431
Hours of Operation: Mon-Fri 9am-5pm
Email: sales@fibreglass.com